Stagroma
 

Scalability

YMC simplifies the process of scaling-up by offering more than 16 fully scalable selectivities with a particle size range from 3 to 150 µm which maintain the integrity of the separation throughout the entire scale-up.

Seamless Scale Up
Reproducible through all Particle Sizes

Drug Discovery ->
Scale-Up -> Production

Uracil
Benzamide
Benzyl alcohol
Phenol
Acetophenone
Methylbenzoate
Anisole
Toluene

Column: ODS-AQ 12 nm, 6.0 x250 mm i.d.
Mobile Phase: acetonitrile/water (50/50)
Flow rate: 2.8 ml/min


 

Media Size Matched to Column Size

For hydrodynamic reasons, smaller media perform best in smaller diameter columns, and larger diameter media excel in larger diameter columns. In general, 3 µm media performs best in columns up to 10 mm i.d.; 5 µm media is best suited to columns up to 30 mm i.d.; and 10 µm media is appropriate for columns up to
50 mm i.d. Columns of 100 mm i.d. or larger are generally packed with particles of 15 µm or larger in diameter.

Large particle media is available in smaller diameter columns (scout columns) to facilitate method development, scale-up and process monitoring. Scale-up Kits are also available.

 

 


© 2010 Stagroma AG
Share |